LAB · QUALITY CONTROL · LEVEY-JENNINGS
QC dashboard · CBC Haemoglobin · Level 1 control
Lot: BIO-CBC-NORMAL-2026-04 · expires 2026-08-15 · mean 14.2 · SD 0.18 · Westgard rules active
CV %
1.27%
Bias %
0.6%
Sigma metric
5.4σ
14-day breach count
1
Levey-Jennings · last 14 days
Westgard rule violations · last 30 days
| Date | Test | Rule | Action taken | By |
|---|---|---|---|---|
| Apr 29 | HGB | 1-3s | Re-run + recalibrate | Naveen V. |
| Apr 22 | HGB | 1-2s | Acknowledged · within tolerance | Naveen V. |
| Apr 14 | WBC | 2-2s | Re-run · accepted | Geetha M. |
| Apr 09 | PLT | R-4s | Lot change verified | Naveen V. |
EQAS / NABL standing
Last EQAS panel (CMC Vellore)Apr 2026 cycle 4 · score 96/100 · A grade
NABL accreditationValid till 2027-08-15
Last NABL audit2026-04-22 · 2 minor obs · CAR closed
ISO 15189 internal auditDue 2026-08-01
CBC · A
Biochem · A
Coag · A
Microbiol · B (subscription pending)
Microbiology EQAS subscription renewal: ₹48,000 · approver: AVP